I. Introduction
CEC201 series dual-channel gas condenser is a kind of compressed high efficiency moisture removing, used in gas pretreatment system of gas analysis system. It can protect moisture and corresponding corrosion to the gas analyser and guarantee the lifespan of gas analysers.
II. Specifications
|
Items |
Descriptions |
|
Working ambient temperature |
5℃~50℃ |
|
Inlet gas temperature |
≤180℃ |
|
Inlet gas dew point |
≤80℃(@ 25℃) |
|
Outlet gas dew point |
≤4℃(@ 25℃) |
|
Condensing control temperature |
Factory set at 4.0℃ |
|
Temperature controller fluctuation |
3.0℃±1.0℃(@ 25℃) |
|
Condensing method |
fluoride-free compressor |
|
Gas condensing pipe |
A+B glass heat exchange pipes, diameter * heightΦ40mm*150mm |
|
Inlet gas pressure |
≤3bar |
|
Allowed gas differential pressure range |
≤3mbar(gas flow 2NL/min) |
|
Gas capacity |
250NL/min |
|
Trouble alarm |
Low temperature alarm, deviation alarm, power off alarm; alarm status dry contact output, capacity 250Vac/1A |
|
Peristaltic pump *2 pcs |
5r/min, discharge capacity 0.3L/h, left IN and right OUT |
|
withstand voltage testing standard |
AC1500V |
|
Leak current |
≤3.5mA |
|
Inlet gas connector |
OD6mm (OD8mm optional) |
|
Water drainage port |
OD6mm (OD8mm optional) |
|
Warming-up time |
≤30min |
|
Installation method |
Wall-mounted plate installation |
|
Power supply |
220VAC 50Hz |
|
Power supply consumption |
400W |
|
Cabinet size(W×H×D) |
310*280*290mm |
|
Cabinet color |
RAL7035 grey |
|
Weight |
18kg |
|
4-20mA output |
optional |
II. Specifications
|
Items |
Descriptions |
|
Working ambient temperature |
5℃~50℃ |
|
Inlet gas temperature |
≤180℃ |
|
Inlet gas dew point |
≤80℃(@ 25℃) |
|
Outlet gas dew point |
≤4℃(@ 25℃) |
|
Condensing control temperature |
Factory set at 4.0℃ |
|
Temperature controller fluctuation |
3.0℃±1.0℃(@ 25℃) |
|
Condensing method |
fluoride-free compressor |
|
Gas condensing pipe |
A+B glass heat exchange pipes, diameter * heightΦ40mm*150mm |
|
Inlet gas pressure |
≤3bar |
|
Allowed gas differential pressure range |
≤3mbar(gas flow 2NL/min) |
|
Gas capacity |
250NL/min |
|
Trouble alarm |
Low temperature alarm, deviation alarm, power off alarm; alarm status dry contact output, capacity 250Vac/1A |
|
Peristaltic pump *2 pcs |
5r/min, discharge capacity 0.3L/h, left IN and right OUT |
|
withstand voltage testing standard |
AC1500V |
|
Leak current |
≤3.5mA |
|
Inlet gas connector |
OD6mm (OD8mm optional) |
|
Water drainage port |
OD6mm (OD8mm optional) |
|
Warming-up time |
≤30min |
|
Installation method |
Wall-mounted plate installation |
|
Power supply |
220VAC 50Hz |
|
Power supply consumption |
400W |
|
Cabinet size(W×H×D) |
310*280*290mm |
|
Cabinet color |
RAL7035 grey |
|
Weight |
18kg |
|
4-20mA output |
optional |
Biomass syngas analyzer CH4/CO2/CO/H2/O2/CxHy TY-6331
Biomass syngas analyzer CH4/CO2/CO/H2/O2/CxHy TY-6331 can measure CH4,CO2,CO,CnHm,O2 and H2 at the same time in one unit, and calculating heating value in MJ/M3 or Kcal/m3.