I. Introduction
Portable biomass syngas multi-gas Analyzer TY-6331P adopts the imported gas sensor measurement using electrochemical sensor of oxygen, hydrogen measurement using MEMS TCD thermal conductivity sensor, methane and carbon dioxide, carbon monoxide, measured using non dispersive infrared gas sensor NDIR principle (selected according to user needs).
II. Features
• Using 5'' TFT LCD touch screen
• Measurement value display/real-time curve display
• Aluminum enclosure with smart dimensions (270*215*136mm, 3kgs)
• Data record, interval time adjustment, USB output to display history data
• Power intelligent management protect gas analyzer working at low battery
• Calibration menu and alarm setting with password protection
• One set back to factory setting
III. Specifications
Measurement |
CO, CO2, CH4, CnHm, O2, H2 and gas calorific value calculation |
Calculation |
High Heating Value or Low Heating Value in MJ/m3 or kcal/m3 |
Technology |
CO, CO2, CH4, CnHm: proprietary dual beam NDIR detectors O2: industrial electrochemical cell H2: MEMS thermal conductivity detector |
Ranges |
CO/ CO2:/CH4 / H2: 0-100%, CnHm : 0-10%, O2: 0-25% Other ranges customizable on request without price increase |
Resolution |
0,01% |
Accuracy |
CO/CO2/CnHm/CH4:±2%FS;H2/O2:±3%FS |
Repeatability |
≤ 1% |
Gas conditions at analyzer inlet |
Flow:0.7L-1.2/min, internal gas sampling pump Pressure:2kPa≤ inlet pressure≤ 50kPa Quality: No dust, moisture, tar (external safety filter and optional gas washing unit) |
Operating conditions |
Tamb : -5°C to 50°C / Pamb : 86 to 108 kPa / RH : 0-90% non-condensing |
Response time (T90) |
≤ 15 sec (NDIR) |
Communication interface |
USB output; RS232/RS485 |
Power supply |
External 220 VAC-50Hz Internal with lithium battery and charger |
Dimensions and weight |
Gas analyser :270(L) x 215 (H) x 136 (W) mm / ± 3kg |
IV.
Applications
III. Specifications
Measurement |
CO, CO2, CH4, CnHm, O2, H2 and gas calorific value calculation |
Calculation |
High Heating Value or Low Heating Value in MJ/m3 or kcal/m3 |
Technology |
CO, CO2, CH4, CnHm: proprietary dual beam NDIR detectors O2: industrial electrochemical cell H2: MEMS thermal conductivity detector |
Ranges |
CO/ CO2:/CH4 / H2: 0-100%, CnHm : 0-10%, O2: 0-25% Other ranges customizable on request without price increase |
Resolution |
0,01% |
Accuracy |
CO/CO2/CnHm/CH4:±2%FS;H2/O2:±3%FS |
Repeatability |
≤ 1% |
Gas conditions at analyzer inlet |
Flow:0.7L-1.2/min, internal gas sampling pump Pressure:2kPa≤ inlet pressure≤ 50kPa Quality: No dust, moisture, tar (external safety filter and optional gas washing unit) |
Operating conditions |
Tamb : -5°C to 50°C / Pamb : 86 to 108 kPa / RH : 0-90% non-condensing |
Response time (T90) |
≤ 15 sec (NDIR) |
Communication interface |
USB output; RS232/RS485 |
Power supply |
External 220 VAC-50Hz Internal with lithium battery and charger |
Dimensions and weight |
Gas analyser :270(L) x 215 (H) x 136 (W) mm / ± 3kg |
IV.