1.Sulfur hexafluoride fast gas analyzer uses advanced infrared absorption technology NDIR
2, can detect multi-component gas, up to 8 kinds of gas, can also be configured according to user needs
3. With automatic pump-back function, the user can choose whether to return the measured gas pump to the circuit breaker.
4. With pipeline vacuum dehumidification function, make the test data more real and accurate
5. Optional 3 measurement cycles, each cycle takes only 1 minute to complete
6, automatic sampling, continuous measurement of 3 cycles
7. Accurate and precise: bring laboratory-level instruments to the scene
8.Sulfur hexafluoride fast gas analyzer optional leak detection function according to user needs
9.Add CMS online detection function according to user needs
|
SF6 (purity) range |
0-100% |
|
Error |
≤ ± 0.1% (90 ~ 100%) |
|
H20 (micro water) range |
-20 ℃ ~ -60 ℃ |
|
Error |
≤ ± 0.5 ℃ |
|
SO2 (sulfur dioxide) Range |
0-150 / 500ppm (Dual range automatic switching) Measurement value ≤10ppm, error ≤ ± 0.4ppm Measurement value ≥10ppm, error ≤ ± 2% |
|
H2S (hydrogen sulfide) Range |
0-100ppm Measurement value ≤10ppm, error ≤ ± 0.5ppm Measurement value ≥10ppm, error ≤ ± 5% |
|
CO (carbon monoxide) range |
0-1000ppm Measurement value ≤50ppm, error ≤ ± 5ppm Measurement value ≥50ppm, error ≤ ± 5% |
|
HF (hydrogen fluoride) Range |
0-200ppm Error: ± 5% FS |
|
CF4 (carbon tetrafluoride) range |
0 -4000ppm error: ± 2% FS |
|
Range |
0-65% Error: ± 1% FS |
|
Air range |
0-50% Error: ± 1% FS |
III. Specifications
|
SF6 (purity) range |
0-100% |
|
Error |
≤ ± 0.1% (90 ~ 100%) |
|
H20 (micro water) range |
-20 ℃ ~ -60 ℃ |
|
Error |
≤ ± 0.5 ℃ |
|
SO2 (sulfur dioxide) Range |
0-150 / 500ppm (Dual range automatic switching) Measurement value ≤10ppm, error ≤ ± 0.4ppm Measurement value ≥10ppm, error ≤ ± 2% |
|
H2S (hydrogen sulfide) Range |
0-100ppm Measurement value ≤10ppm, error ≤ ± 0.5ppm Measurement value ≥10ppm, error ≤ ± 5% |
|
CO (carbon monoxide) range |
0-1000ppm Measurement value ≤50ppm, error ≤ ± 5ppm Measurement value ≥50ppm, error ≤ ± 5% |
|
HF (hydrogen fluoride) Range |
0-200ppm Error: ± 5% FS |
|
CF4 (carbon tetrafluoride) range |
0 -4000ppm error: ± 2% FS |
|
Range |
0-65% Error: ± 1% FS |
|
Air range |
0-50% Error: ± 1% FS |


Infrared natural gas calorimeter analyzer TY-6340
The online infrared natural gas calorimeter analyzer TY-6340 uses the most advanced NDIR non-dispersive infrared technology in the world. It is mainly used to measure the volume concentration and calorific value of gas such as CH4 and CnHm in natural gas. The online infrared natural gas calorimeter
Portable biomass syngas multi-gas Analyzer TY-6331P
syngas analyzer to measure CH4/CO2/CO/H2/CxHy/O2 and calorific value for biomass gasification/coal gasification/pyrolysis process gas analysis monitoring.