I. Introduction
The Infrared landfill biogas analyzer TY-6321 uses the most advanced NDIR non-spectral infrared technology and ECD electrochemical technology. It is mainly used to measure CO, CO2, CH4, H2, O2, H2S six gas concentrations in biogas and various anaerobic fermentation processes and experiments.
Measurement
CH4, CO2, O2,
H2S (extendable to CO,H2)
Technology
CO, CH4 , CO2:
proprietary dual beam NDIR detectors
O2 ,H2S: industrial
electrochemical cell
H2: TCD
Ranges
CH4: 0-100%, CO2 : 0-50%, O2:
0-25%, H2S:0-9999ppm, CO:0-2000ppm, H2: 0-1000ppm
Other ranges customizable on request without
price increase
Resolution
0,01% for CH4, CO2, O2; 1ppm for H2S, CO, H2
Accuracy
CO /CO2/CH4:<=2%FS;O2 /H2S:1ppm
Repeatability
≤1%FS
Gas conditions at analyzer inlet
Flow:0.7L-1.2/min
Pressure:2kPa≤ inlet
pressure≤ 50kPa
Quality: No dust,
moisture, tar (external safety filter and optional gas washing unit)
Operating conditions
Tamb : -5°C to 50°C / Pamb : 86 to 108 kPa / RH :
0-90% non-condensing
Response time (T90)
≤ 15 sec
(NDIR)
Communication interface
4-20mA
output; RS485
Power supply
External 110V-220 VAC-50Hz
Dimensions and weight
Gas
analyser :482(L) x 134 (H) x 320 (W) mm / 5kg
II. Features
1. Standardized 3U chassis design, aluminum alloy brushed panel chassis, stylish appearance, strong sense of technology, light instrument weight.
2. Imported NDIR non-spectral infrared sensor and ECD electrochemical gas sensor.
3. The instrument measures the volume concentration of CO, CO2, CH4, H2, O2, and H2S in the sample gas. Add the over-range limit enable function to ensure the measurement accuracy within the range, the over-range can be measured and provide a reference for the measured value.
4. The internal gas interference of the instrument is automatically corrected to ensure no interference in the measurement results.
5. The instrument has the function of intelligent determination of user-calibrated ventilation concentration to prevent misoperation by the user and cause the instrument to be unusable.
6. The instrument has fast response time and can realize 7 * 24 * 365 continuous monitoring.
7. The instrument automatically samples, analyzes, detects, and measures quickly and uses no consumables.
8. The equipment is easy to operate and can be used after the personnel are trained on site.
9. The equipment is suitable for continuous monitoring of industrial sites and sampling and analysis of airbags in laboratories.
10. Chinese and English software operating systems can be switched freely.
11. The instrument software is powerful, with functions such as automatic data storage, query, deletion, and USB export. The history data storage site name can be customized (support Chinese input), and the measurement data storage interval can be customized.
12. With RS-485 digital output interface
13. Modular product design, software version can be upgraded remotely.
III.
Specifications
IV. Applications
Landfill
Sewage treatment
Anaerobic digestion
CDM projects
Medium and large size biogas power plant
Measurement |
CH4, CO2, O2, H2S (extendable to CO,H2) |
Technology |
CO, CH4 , CO2: proprietary dual beam NDIR detectors O2 ,H2S: industrial electrochemical cell H2: TCD |
Ranges |
CH4: 0-100%, CO2 : 0-50%, O2: 0-25%, H2S:0-9999ppm, CO:0-2000ppm, H2: 0-1000ppm Other ranges customizable on request without price increase |
Resolution |
0,01% for CH4, CO2, O2; 1ppm for H2S, CO, H2 |
Accuracy |
CO /CO2/CH4:<=2%FS;O2 /H2S:1ppm |
Repeatability |
≤1%FS |
Gas conditions at analyzer inlet |
Flow:0.7L-1.2/min Pressure:2kPa≤ inlet pressure≤ 50kPa Quality: No dust, moisture, tar (external safety filter and optional gas washing unit) |
Operating conditions |
Tamb : -5°C to 50°C / Pamb : 86 to 108 kPa / RH : 0-90% non-condensing |
Response time (T90) |
≤ 15 sec (NDIR) |
Communication interface |
4-20mA output; RS485 |
Power supply |
External 110V-220 VAC-50Hz |
Dimensions and weight |
Gas analyser :482(L) x 134 (H) x 320 (W) mm / 5kg |